COMBINATION pH ELECTRODE ERH-11S


ERH-11S combination pH electrode has a specific construction of the electrolytical junction made in a shape of sliding, glass cut. It is a glass sleeve precisely adjusted to the electrode's body, covering the junction, which protects it from clogging. Accurate construction enables stable outflow of an excess electrolyte and good contact with it.
ERH-11S is designed for measurements of substances with content of water higher than 5%

May be used for measurements in:


Glass membrane responds well to the pH value of the measured solutions. The side tubus facilitates refilling or replacement of the electrolyte. The electrode construction minimalises the influence of the static charge occuring frequently in case of measurements in solutions with low ion content.


Technical Data
Measuring Range 0 ÷ 14 pH
Temperature range 0 ÷ 60 °C
for short period usage 80 °C
Resistivity of the membrane 100 ÷ 200 MΩ at 20 °C
Resistivity of the diaphragms < 5 kΩ
Zero point 0 ± 30 mV
Reference electrode Ag/AgCl
Reference solution 3.0 M KCl +AgCl
Body diameter 12.0 mm ±0.5 mm
Body length (without cable socket) 120 mm ±5 mm
Minimal depth of immersing 30 mm
Maximal depth of immersing 105 mm
Membrane material glass
Membrane shape round, bulb
Diaphragm glass, sleeve
Cable length about 1 m
Connector BNC-50